TB 9-5985-314-35
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
11. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 11 and
12 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
a.
Use Appendix B to
determine TI
parameters for directivity.
NOTE
Clean all connectors with alcohol before proceeding with b
below.
b. Connect equipment as
shown in
figure 1 and
allow 1 hour
for equipment warm-up.
NOTE
Do not connect TI into equipment setup at this time.
c. Determine and record a minimum of 10 evenly spaced test frequencies over the TI
a. Performance Check
and adjust RF OUTPUT controls for +6 dBm. Connect N-short to TI DEVICE UNDER TEST
port.