TB 9-6625-047-24
Table 3. Range Accuracy Continued
Test instrument
Calibrator
Adjustments
RANGE
Meter
Initial input
Error display indication
(fig. 1)
(VOLTS)
(R)
indication
Voltage
(+%)
.03
3
30
mV
1.0
kHz
1.0
---
.03
--
30
mV
1.0
kHz
N/A1
---
.1
1
100
mV
10
kHz
1.0
---
.1
--
90
mV
1.0
kHz
N/A1
---
.3
3
.3 V
1.0
kHz
1.0
---
.3
--
.3 V
1.0
kHz
N/A1
---
1
1
1.0 V
1.0
kHz
1.0
---
1
--
0.9 V
1.0
kHz
N/A1
---
3
3
3
V
1.0
kHz
1.0
---
3
--
3
V
1.0
kHz
N/A1
---
10
1
10
V
10
Hz
5.0
---
10
1
10
V
1.0
kHz
1.0
---
10
1
10
V
200
kHz
1.0
---
30
3
30
V
10
Hz
5.0
---
30
3
30
V
1.0
kHz
1.0
---
30
3
30
V
200
kHz
1.0
---
100
1
100
V
10
Hz
5.0
---
100
1.0
kHz
---
.8
80
V
1.25
100
200
kHz
---
.5
50
V
2.0
300
20
Hz
---
2
200
V
6.25
300
1.0
kHz
---
3
300
V
1.0
300
200
kHz
---
1
100
V
3.0
1Record resulting TI indication for use in (5) below.
Figure 1. Test instrument (top view) - adjustment locations.