TB 9-6625-047-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of AC Voltmeter Hewlett-Packard, Model 400EL-02, ME-459/U (Hewlett-
Packard, Model 400EL), and ME-465/U (Hewlett Packard, Model 400E).
The
manufacturers' manuals were used as the prime data sources in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time
required for this
calibration is
approximately 1.5
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument
parameters
Performance specifications
Ac voltage
Range: 1.0 mV to 300 V
Accuracy: (% of FS + % of reading)
Frequency
Range (V)1 2
.001 to 300............................2.5 + 2.5
10
to
40 Hz
.001 to 300............................1.0 + 0.0
40 Hz to 500 kHz
2.0 MHz .001.......................................2.5 + 2.5
500 kHz to
.003 to 300............................1.0 + 0.0
4.0 MHz .001.......................................2.5 + 2.5
2.0
to
.003 to 300............................1.5 + 1.5
10 MHz .001.......................................N/A
4.0
to
.003 to 30..............................2.5 + 2.53
100 to 300.............................N/A
See footnotes at end of table.