TB 9-6625-047-24
Table 1. Calibration Description - Continued
Test instrument
parameters
Performance specifications
Ac to dc converter
Range: .001 to 300 V
output
Accuracy: (% of FS + % of reading)4
Range (V)1
.001 to 300 .................................... 2.5 +
2.5
1.0
to 20
Hz
.001 to 300 .................................... 1.0 +
1.0
20
to 100
Hz
.001 ............................................... 0.4 +
0.1
100 Hz to 500
kHz
.003 to 300 .................................... 0.25 +
0.25
.001 ............................................... 1.0 +
1.0
500 kHz to
1.0 MHz
.003 to 300 .................................... 0.5 +
0.5
.001 to 300 .................................... 2.5 +
2.5
1.0
to
4.0 MHz
.001 ............................................... N/A
4.0
to
10
MHz
.003 to 30 ...................................... 2.5 +
2.53
100 to 300 ..................................... N/A
1Accuracy
applies to 1/3 FS to FS (.001 V range only).
2Accuracy
for the 3 V range at 1/3 FS and below; 6 to 10 MHz is 3.75 + 3.75.
3For serial numbers 1208A22144, 1208A22173, 1208A22184, 1208A22221, 1208A22259, and above for ME-465/U (model
400E) and 1208A22209, 1208A22221, 1208A22227, 1208A22231, 1208A22258, and above for ME-459/U (model 400EL): .01 to
3 V range, 4.0 to 10 MHz accuracy is 3.0 + 2.0; 10 to 30 V range, 4.0 to 10 MHz accuracy is 3.5 + 3.5.
4Verified only at 10 mV, 1 kHz.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
calibration procedure.