TB 9-6625-086-24
a. Performance Check
NOTE
Do not perform the power supply check if all other parameters
are within tolerance.
(1) Connect multimeter to pin W of terminal board XA2 (fig. 4) and ground.
(2) If the multimeter does not indicate between -17.98 and -18.02 volts dc, perform
b below.
(3) Vary the output of autotransformer between 105 and 125 volts ac.
The
multimeter indication will remain within the limits specified in (2) above.
(4) Adjust the autotransformer for a 115 volt ac output.
a. Deenergize and disconnect all equipment and replace TI protective cover.
b. Annotate and affix DA label/form in accordance with TB 750-25.