TB 9-6625-091-35
SECTION III
PRELIMINARY OPERATIONS
6. Preliminary Instructions
a. The
instructions outlined in this
section are preparatory to the
calibration process.
Personnel should become familiar with the entire bulletin before beginning the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
WARNING
HIGH VOLTAGE is used during the performance of this
calibration. DEATH ON CONTACT may result if personnel
fail to observe safety precautions.
7. Equipment Setup
b.
Connect extension (B4) to TI and
insert into
vertical compartment of
oscilloscope.
c. Energize equipment and allow at least 15 minutes for warm-up.
d. Turn TI VOLTS/CM switch to .05 and AC-DC switch to AC.
e. Using TI VERTICAL POSITION control, position trace on crt graticule centerline.
If trace cannot be positioned to centerline, set VERTICAL POSITION control to midrange
and adjust R6821 (fig. 1) until trace is positioned on crt graticule centerline.
f. Turn TI VARIABLE control throughout its range and observe that trace does not
move vertically. If necessary, adjust DC BAL control (front panel) until trace stabilizes.
g. Turn TI VARIABLE control to CALIBRATED.
SECTION IV
NOTE
Unless otherwise specified, verify the results of each test and
take corrective action whenever the test requirement is not met
4