TB 9-6625-091-35
Figure 2. Plug-in - top view
a. Performance Check
(1) Connect square-wave generator (A3) hi amplitude output to TI INPUT, using
cable and termination supplied with square-wave generator, standardizer (B5), adapter
(B1), and if required, adapter (B2).
(2) Turn TI VOLTS/CM switch to .05.
(3) Adjust square-wave generator amplitude for 4 centimeters of vertical deflection
(4) Adjust standardizer for optimum square-wave display on crt. If square waves do
not have flat tops, adjust C5671 (fig. 1) to midrange and readjust standardizer for optimum
square waves.
(5) Repeat technique of (2) and (3) above for TI VOLTS/CM switch settings listed in
table 6. Adjust for optimum square-wave display. Do not
readjust standardizer.
7