Oscillator
SECTION IV. CALIBRATION PROCESS FOR ESI, MODEL 250DE

Calibration Procedure For Capacitance, Inductance, And Resistance Test Set An/Urm-90 (Zm-30/U) And Universal Impedance Bridge Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
. . .
TB 9-6625-1014-24
(4) Adjust autotransformer output down to 105 V ac than up to 125 V ac, while
observing frequency counter. Indicated frequency will remain between 990 and 1010 Hz.
(5) Adjust autotransformer output to 115 V ac.
Figure 2. Oscillator - equipment setup.
b. Adjustments. No adjustments can be made.
12. Capacitance
a. Performance Check
(1) Connect 0.l F capacitance standard to C binding posts.
(2) Position controls as listed in (a) through (f) below:
CIRCUIT SELECTOR switch to D-QX.01.
(a)
AMP RESPONSE switch to PEAK.
(b)
DETECTOR switch to INTERNAL.
(c)
GENERATOR switch to INTERNAL.
(d)
LRC dial MULTIPLIER switch to .01 F.
(e)
LRC DIAL to 10.00.
(f)
9