TB 9-6625-1097-35
(5) Check frequency and voltage at 100, 200, 700, and 900 kHz. Frequency counter
will indicate within 1% of selected frequency and true rms voltmeter will indicate between
90 and 125 mV. If necessary, adjust ADJ 1MC, C112, and C130 for optimum performance.
9. Frequency
Response and Output
Voltage
a. Performance Check
(1) Connect equipment as shown in figure 1, CONNECTION B.
(2) Press 1.00 FREQUENCY DIGITS pushbuttons and X1 KC
MULTIPLIER pushbutton.
(3) Adjust VERNIER control to CAL.
(4) Adjust AMPLITUDE control for a 1.00 V ac indication on multimeter.
(5) Press 5.00 FREQUENCY DIGITS pushbuttons. Multimeter will indicate
between .98 and 1.02 V ac.
(6) Press MULTIPLIER and FREQUENCY DIGITS pushbuttons as listed in table 4.
Multimeter will indicate between .98 and 1.02 V ac.
Table 4. Frequency Response -
Low Frequency
Test instrument pushbutton settings
Frequency digits
Multiplier
1st
2d
3d
X1 KC
9
0
0
X100 CPS
1
0
0
X100 CPS
5
0
0
X100 CPS
9
0
0
X10 CPS
2
0
0
X10 CPS
5
0
0
X10 CPS
9
0
0