TB 9-6625-1098-24
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
300 V range
2%, 30 Hz to 300 kHz
5%, 20 Hz to 500 kHz
1 mV to 30 V range
2%, 10 Hz to 1 MHz
5%, 5 Hz to 3 MHz
100 to 300 V range
2%, 10 Hz to 300 kHz
5%, 5 Hz to 500 kHz
High pass filter
3 dB down at 400 Hz; 60 Hz rejection >40 dB
(Hewlett-Packard Models 333A and 334A)
Low pass filter
3 dB down at 30 kHz; -23.6 dB at 60 kHz
(Hewlett-Packard Model 334A option 002)
<25 V rms with input shielded and terminated with
Residual noise
S
600
et level sensitivity
0.3 rms for 100% set level (up to 300 V may be
attenuated to set level reference)
5%, 5 Hz to 300 kHz
10%, 300 to 600 kHz
5 to 20 Hz, >+1 dB
20 Hz to 20 kHz > 0.6 dB
20 to 100 kHz .-1 dB
100 to 300 kHz >-2 dB
300 to 600 kHz >-3 dB
Residual distortion1
>-70 dB, 5 Hz to 200 kHz
>-64 dB, 200 to 600 kHz
Automatic null
0 to +3 dB from manual null (5 to 100 Hz)
(Hewlett-Packard Models 333A and 334A)
0 to +1.5 dB from manual null (100 Hz to
600 kHz)
>80 dB
1