APPENDIX A. SAMPLE CALIBRATION TEST REPORT
REPORT OF CALIBRATION FOR DECADE CAPACITOR IET MODEL HACS-Z-B-3E-1F

Calibration Procedure For Decade Capacitor, Model Hacs-Z-B-6E-1Pf And Hacs-Z-B-3E-1Uf Page Navigation
  11    12    13    14    15  16  17    18  
img
TB 9-6625-113-24
APPENDIX A
Measured Capacitance 1000 pF Steps (1 kHz).
Decade position
Min (pF)
Measured
Max (pF)
Uncertainty (%)
1
998.00
1002.0
0.0216
2
1997.0
2003.0
0.0208
3
2996.0
3004.0
0.0205
4
3995.0
4005.0
0.0204
5
4994.0
5006.0
0.0203
6
5993.0
6007.0
0.0203
7
6992.0
7008.0
0.0202
8
7991.0
8009.0
0.0202
9
8990.0
9010.0
0.0202
10
9989.0
10011
0.0202
Measured Capacitance 0.01F Steps (1 kHz).
Decade position
Min (nF)
Measured
Max (nF)
Uncertainty (%)
1
9.9890
10.001
0.0202
2
19.979
20.021
0.0202
3
29.969
30.031
0.0202
4
39.959
40.041
0.0202
5
49.949
50.051
0.0202
6
59.939
60.061
0.0202
7
69.929
70.071
0.0202
8
79.919
80.081
0.0202
9
89.909
90.091
0.0202
10
99.899
100.10
0.0202
Measured Capacitance 0.1F Steps (1 kHz).
Decade position
Min (nF)
Measured
Max (nF)
Uncertainty (%)
1
99.899
100.10
0.0202
2
199.80
200.20
0.0202
3
299.70
300.30
0.0202
4
399.60
400.40
0.0202
5
499.50
500.50
0.0202
6
599.40
600.60
0.0202
7
699.30
700.70
0.0202
8
799.20
800.80
0.0202
9
899.10
900.90
0.0202
10
999.00
1001.0
0.0202
Under the conditions stated, the limits of error are as indicated. This calibration is traceable to, and
compatible with, the National Institute of Standards (NIST) measurements.
Calibration Report No. ____________
Calibrating Technician: _________________
Temperature: __________
Verified by: ____________________________
Date: __________
Page 2 of 2
A-2