Figure 5. Capacitance Measure HACS-Z-B-3E-1uF.
Table 12. Capacitance Accuracy 1 mF Steps (100 Hz).

Calibration Procedure For Decade Capacitor, Model Hacs-Z-B-6E-1Pf And Hacs-Z-B-3E-1Uf Page Navigation
  8    9    10    11    12  13  14    15    16    17    18  
img
. . . . . . .
TB 9-6625-113-24
(6) Repeat technique of (4) and (5) above for remaining TI capacitance values in
table 9. When decade is complete, return switch to the 0 position.
(7) Disconnect impedance measuring system from TI.
(8) Change the impedance measuring system test frequency to 100 Hz, then
perform open and short compensation on impedance measuring system (fig. 4, connections
A and B).
(9) Reconnect impedance measuring system to TI as shown in figure 5.
(10) Set the TI to the first value in table 10.
(11) TI will indicate within the limits of table 10.
(12) Repeat technique of (10) and (11) above for remaining TI capacitance values in
tables 10 through 12. When decade is complete, return switch to the 0 position.
Table 10. Capacitance Accuracy 10 F Steps (100 Hz).
TI
Impedance measuring system
Measured
Min (F)
Max (F)
Decade position
1
9.9900
10.010
2
19.980
20.020
3
29.970
30.030
4
39.960
40.040
5
49.950
50.050
6
59.940
60.060
7
69.930
70.070
8
79.920
80.080
9
89.910
90.090
10
99.900
100.10
Table 11. Capacitance Accuracy 100 F Steps (100 Hz).
TI
Impedance measuring system
Measured
Min (F)
Max (F)
Decade position
1
99.500
100.50
2
199.00
201.00
3
298.50
301.50
4
398.00
402.00
5
497.50
502.50
6
597.00
603.00
7
696.50
703.50
8
796.00
804.00
9
895.50
904.50
10
995.00
1005.0