TB 9-6625-113-24
(6) Repeat technique of (4) and (5) above for remaining TI capacitance values in
table 9. When decade is complete, return switch to the 0 position.
(7) Disconnect impedance measuring system from TI.
perform open and short compensation on impedance measuring system (fig. 4, connections
A and B).
(10) Set the TI to the first value in table 10.
(11) TI will indicate within the limits of table 10.
(12) Repeat technique of (10) and (11) above for remaining TI capacitance values in
tables 10 through 12. When decade is complete, return switch to the 0 position.
TI
Impedance measuring system
Measured
Min (F)
Max (F)
Decade position
1
9.9900
10.010
2
19.980
20.020
3
29.970
30.030
4
39.960
40.040
5
49.950
50.050
6
59.940
60.060
7
69.930
70.070
8
79.920
80.080
9
89.910
90.090
10
99.900
100.10
TI
Impedance measuring system
Measured
Min (F)
Max (F)
Decade position
1
99.500
100.50
2
199.00
201.00
3
298.50
301.50
4
398.00
402.00
5
497.50
502.50
6
597.00
603.00
7
696.50
703.50
8
796.00
804.00
9
895.50
904.50
10
995.00
1005.0
14