TB 9-6625-113-24
NOTE
system may have a negative indication because the zero
capacitance measurement and the value stored during the zero
compensation process are nearly equal. In this situation, the
TI zero capacitance is considered negligible.
(4) Set the TI to the first value in table 9 (0-0-0-0).
Table 9. Capacitance Accuracy 1 F Steps (1 kHz).
TI
Measured
Min (F)
Max (F)
Decade position
0
0.00000 pF
0.50000 pF
1
0.9990
1.0010
2
1.9980
2.0020
3
2.9970
3.0030
4
3.9960
4.0040
5
4.9950
5.0050
6
5.9940
6.0060
7
6.9930
7.0070
8
7.9920
8.0080
9
8.9910
9.0090
10
9.9900
10.010
(5) Impedance measuring system will indicate within the limits of table 9.
13