TB 9-6625-116-24
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Oscilloscope calibrator base (mainframe), Fluke, Model 9500B/3200. The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test Instrument Parameters
Performance Specifications
DC voltage
Range: 1 mV to 200 V into 1 M
1 mV to 5 V into 50
Accuracy: (0.025% + 25 uV)
Square wave voltage
Range: 40 uV to 200 V p-p into 1 M 10 Hz to <10 kHz
Accuracy: 1 mV (0.1% + 10 uV)
Accuracy: < 1 mV (1% + 10 uV)
Range: 1.9 V to 4.8 V @ 1 kHz and 45 kHz
Accuracy: 3.5%
Time marker
Range: 9.0091 nS to 55s
Accuracy: 0.25 ppm
Pulse width
Range: 1 ns to 100 ns
Accuracy: < 5% 200 ps
Range: 10 to 40
Accuracy: 0.5%
Range: 40 to 90
Accuracy: 0.1%
Range: 90 to 150
Accuracy: 0.5%
Range: 50 k to 800 k
Accuracy: 0.5%
Range: 800 k to 1.2 M
Accuracy: 0.1%
Range: 1.2 M to 12 M
Accuracy: 0.5%