Table 3. DC Voltage Accuracy Continued.
Table 5. AC Sinewave Voltage Accuracy

Calibration Procedure For Oscilloscope Calibrator Base (Mainframe) Model 9500B/3200 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . .
TB 9-6625-116-24
(10) Repeat technique of (8) (a) through (d) above for remaining TI voltage outputs
and WAVE FORMs in table 4.
NOTE
Limits in table 4 were calculated using manufacturer's formula
(0.5 x Vp-p x 0.999917). 0.999917 (83 ppm) compensates for
the finite transition time of the square wave and the resultant
fall in its RMS value.
Table 4. AC Squarewave Voltage Accuracy
TI
Multimeter
Output (V p-p)
Wave Form
Min
Measured
Max
0.006
2.99175 mV
3.00775 mV
Positive Square
0.006
Negative Square
2.99175 mV
3.00775 mV
0.006
Symmetrical Square
2.99175 mV
3.00775 mV
0.06
Positive Square
29.9625 mV
30.0325 mV
0.06
29.9625 mV
30.0325 mV
Negative Square
0.06
Symmetrical Square
29.9625 mV
30.0325 mV
0.6
0.299670 V
0.300280 V
Positive Square
0.6
Negative Square
0.299670 V
0.300280 V
0.6
Symmetrical Square
0.299670 V
0.300280 V
6
Positive Square
2.99675 V
3.00276 V
6
2.99675 V
3.00276 V
Negative Square
6
Symmetrical Square
2.99675 V
3.00276 V
601
Positive Square
29.9675 V
30.0275 V
601
Negative Square
29.9675 V
30.0275 V
601
Symmetrical Square
29.9675 V
30.0275 V
To avoid "load <50k
detected" warning message, set OUTPUT to OFF, change waveforms, then set
1
OUTPUT to ON.
(11)Set TI OUTPUT to OFF.
(12) Disconnect active head at multimeter, insert 50
feedthrough termination and
reconnect to multimeter.
(13) Set TI as listed in (a) through (m) below:
(a) Major function,
(sine)
(b) [
/ 1.0] (select 1.0)
(c) [CHANNEL SELECT]
]
(d) [LOAD 50
7