Table 4. AC Squarewave Voltage Accuracy
Table 6. Time Marker Accuracy

Calibration Procedure For Oscilloscope Calibrator Base (Mainframe) Model 9500B/3200 Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . . . .
TB 9-6625-116-24
(e) [SIGNAL CH 1]
(f) [EXIT]
(g)
(if necessary) to select AMPLITUDE
(h) Keypad Entry, 4.8
(i) [V]
(j)
(if necessary) to select FREQUENCY
(k) Keypad Entry, 1
(l) [kHz]
(m) OUTPUT, ON
(14) Multimeter will indicate within the limits of table 5. If not, perform b below.
(15) Repeat technique of (13) (g) through (l) above for remaining TI voltages and
frequencies in table 5.
Table 5. AC Sinewave Voltage Accuracy
TI
Multimeter
Output (V p-p)
Frequency (kHz)
Min (V rms)
Measured
Max (V rms)
4.8000
1
1.63770
1.75650
4.8000
45
1.63770
1.75650
1.9000
1
0.648200
0.695300
1.9000
45
0.648200
0.695300
(16) Set TI OUTPUT to OFF and disconnect equipment set-up.
b. Adjustments. Perform entire alignment procedure in paragraph 11 below.
9. Time Marker and Pulse Width Accuracy
a. Performance Check
NOTE
To minimize errors associated with precise trigger levels,
hysteresis compensation should be performed on frequency
counter before measuring TI pulse width. Ensure that nothing
is connected to frequency counter. Press AUX MENU, with the
SELECT/SET key, select CAL HYST then press ENTER.
(1) Connect time/frequency workstation 10MHz output to frequency counter
REFERENCE IN.
(2) Connect TI CH 1 output to frequency counter INPUT A through active head.