Table 7. Pulse Width Accuracy
Table 9. Special Cal DAC Characterization

Calibration Procedure For Oscilloscope Calibrator Base (Mainframe) Model 9500B/3200 Page Navigation
  5    6    7    8    9  10  11    12    13    14    15  
img
. . . . . . . .
TB 9-6625-116-24
(4) TI will indicate within the limits of table 8. If not, perform b below.
(5) Set resistance standard to 1.00000 M .
].
(6) Change TI load setting to 1 M
by pressing [LOAD 1 M
(7) TI will indicate within the limits of table 8. If not, perform b below.
Table 8. Load Resistance Measurement Accuracy
Resistance Standard
TI
Min
Measured
Max
50.010
49.960
50.060
1.00000 M
0.9990 M
1.0010 M
(8) Set TI OUTPUT to OFF and disconnect equipment set-up.
b. Adjustments. Perform entire alignment procedure in paragraph 11 below.
11. Alignment
a. Unsecuring TI for Alignment
(1) Set rear panel CAL switch to ENABLE.
(2) Set TI as listed in (a) through (c) below:
(a) Major function, MODE
(b) [CALIB]
(c) Keypad Entry, 2
(default password)
3
5
7
b. Special Cal
NOTE
SPECIAL CAL must be completed before performing BASE
CAL. There is no connection to external standards during the
SPECIAL CAL routine (except for ADJUST FREQ).  Once
SPECIAL CAL is completed, the entire BASE CAL must be
completed because numerous internal adjustments were made.
(1) Set TI as listed in (a) and (b) below:
(a) [SPECIAL CAL]
(b) [Chse DAC] (first characterization in table 9)
(2) Chse - DAC will be highlighted and all other Special Cal options will blank.
DAC characterization is complete when the single highlighted key is replaced by complete
softkey menu display.
(3) Repeat technique of (1) (b) and (2) above for remaining Special Cal
characterizations in table 9.