Table 11. Sine wave P-P to RMS Conversion (1 kHz and 45 kHz)

Calibration Procedure For Oscilloscope Calibrator Base (Mainframe) Model 9500B/3200 Page Navigation
  10    11    12    13    14  15
img
. . .
TB 9-6625-116-24
(f) [DEFAULT TARGETS] (select default targets, not saved targets)
(g) [TARGET 1]
(h) OUTPUT, ON
(28) Set resistance standard to TI TARGET RES value.
NOTE
When applied resistance differs from the TARGET RES value,
the TARGET RES value can be changed to the applied
resistance (i.e. 50.010  or 19.996 M ).
(29) Press [ACCEPT CALIB].
(30) Press [NEXT TARGET].
(31) Repeat technique of (28) through (30) above for remaining targets. When
TARGET RES value is 19 M , substitute calibrator for resistance standard.
(32) When last target is completed, press MODE.
(33) Press [EXIT].
(34) Set rear panel CAL switch to DISABLE position.
(35) Repeat paragraphs 8 through 10 above.
12. Final Procedure
a. If desired, TI internal base CAL DUE DATE may be updated by setting TI as listed
(1) through (10) below.
(1) Rear panel CAL switch to ENABLE
(2) Major function, MODE
(3) [CALIB]
(4) Keypad Entry, 2
(default password)
3
5
7
(5) [BASE CAL]
(6) Major function, MODE
(7) [CAL DUE DATE]
25
2012 depending on the setting
(8) Enter new cal due date followed by
(i.e 12
of the TI date format).
(9) [EXIT]
(10) Rear panel CAL switch to DISABLE
b. De-energize and disconnect all equipment.
c. Annotate and affix DA label/form in accordance with TB 750-25.
16