TB 9-6625-116-24
(19) Set TI as listed in (a) through (h) below:
(a) Major function, (sine)
(b) [CHANNEL SELECT]
]
(c) [LOAD 50
(d) [SIGNAL CH 1]
(e) [EXIT]
(f) [DEFAULT TARGETS] (select default targets, not saved targets)
(g) [TARGET 1]
(h) OUTPUT, ON
(20) Using cursors or spinwheel, adjust O/P AMPLITUDE until multimeter
indicates the TARGET AMPLITUDE.
NOTE
Use table 11 to convert from sine wave p-p voltage to RMS
voltage.
Table 11. Sine wave P-P to RMS Conversion (1 kHz and 45 kHz)
Sine wave V p-p
V rms
1.6
0.565685
1.9
0.671751
2.6
0.919239
4.8
1.697056
(21) Press [ACCEPT CALIB].
(22) Press [NEXT TARGET].
(23) Repeat technique of (20) through (22) above for remaining targets.
(24) Set TI OUTPUT to OFF and press [EXIT] when last target is completed.
(25) Disconnect equipment set-up.
(26) Connect TI CH 1 output to resistance standard SENSE HI and SENSE LO
through active head.
(27) Set TI as listed in (a) through (h) below:
(a) Major function, AUX
(b) [CHANNEL SELECT]
]
(c) [LOAD 50
(d) [SIGNAL CH 1]
(e) [EXIT]
15