TB 9-6625-1314-35
Figure 3. Dc and ac current (AN/GSM-256, AN/GSM-259 and AN/TSM-55) - equipment setup.
(3) Set FUNCTION switch to AMPS and RANGE switch to 100 A.
CAUTION
All decade positions of decade resistance standard (A8) must be
set exactly as specified in (4) below and table 9 to assure
maximum current limits of the decades are not exceeded.
Table 9. Dc Current (AN/GSM-256, AN/GSM-259, and AN/TSM-55)
Test instrument
RANGE
Indication
Min
Max
standard setting
switch
001000.00
999.9
0.9969
1.0029
1000 A
000100.00
10 MA
9.999
0.9969
1.0029
000010.00
100 MA
99.99
0.9969
1.0029
17