TB 9-6625-1314-35
(3) Press TI mA function pushbutton and .2 range pushbutton.
(4) Adjust dc voltage standard (A7) and decade resistor (A9) to obtain a 199.9
indication on TI. Ac/dc voltmeter (A3) will indicate between 19.91 and 20.07 mV.
Figure 7. Dc current - equipment setup.
(5) Repeat technique of (2) through (4) above, using settings and indications listed
in table 14. Ac/dc voltmeter will indicate within limits specified.
31