(23) Repeat (22) above with electrical test set frequency set to 400 Hz.
b. Adjustments
(1) Adjust signal source to obtain an indication of 1.9990 V on ac/dc voltmeter.
(2) Adjust 2 VAC (fig. 6) to obtain an indication of 1.999 on TI (R).
(3) Adjust signal source to obtain an indication of 0.19990 V on ac/dc voltmeter.
(4) Adjust .2 VAC (fig. 6) to obtain an indication of 199.9 on TI (R).
(5) Adjust ac power supply on test equipment to obtain an indication of 19.990 V on
(6) Adjust 20 VAC (fig. 6) to obtain an indication of 19.99 on TI (R).
31. Ac Current (Level A)
a. Performance check
Figure 8. Low ac current - equipment setup.
36