SECTION III. CALIBRATION PROCESS

Calibration Procedure For Scopemeter Models 190-202 And 190-202/Ar Page Navigation
1  2    3    4    5    6  
img
. . . . . . . .
TB 9-6625-134-24
Table 1. Calibration Description Continued.
Test instrument parameters
Performance specifications
Oscilloscope - Continued
Time Base
Accuracy: (100 ppm +0.04 div)
Trigger Sensitivity
Range:
Accuracy:
DC to 5 MHz at 2 to 5mV/div
1 division
DC to 5 MHz at > 5 mV/div
0.5 divisions
200 MHz
1 division
250 MHz
2 divisions
Bandwidth
DC to 200 MHz at -3 dB
Meter
DC Voltage
Range:
Accuracy:
500 mV to 1100 V
(0.5% + 5 Counts)
AC Voltage
Range: 500 mV to 1100 V
Range:
Accuracy:
15 to 60 Hz
(1% + 10 Counts)
60 Hz to 1 kHz
(2.5% + 15 Counts)
Ohms
Range:
Accuracy:
500  to 30 M
(0.6% + 5 Counts)
Diode
Range:
Accuracy:
Max Reading 2.8 V
(2% + 5 Counts)
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878; and Secondary Transfer Calibration Standards Set,
AN/GSM-286; AN/GSM-287; or AN/GSM-440.  Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parentheses.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.