TB 9-6625-134-24
Table 6. Oscilloscope Channel A Readout DCV.
Calibrator output
TI
(V)
Range/div (V)
Minimum (V)
Maximum (V)
6m
4.9 m
7.1 m
2m
-6 m
-7.1 m
-4.9 m
15 m
14.3 m
15.7 m
5m
-15 m
-15.7 m
-14.3 m
30 m
29.1 m
30.9m
10 m
-30 m
-30.9 m
-29.1 m
60 m
58.6 m
61.4 m
20 m
-60 m
-61.4 m
-58.6 m
100 m
98 m
102 m
50 m
-100 m
-102 m
-98 m
300 m
291m
309 m
100 m
-300 m
-309 m
-291 m
600 m
586 m
614 m
200 m
-600 m
-614 m
-586 m
1
980 m
1.020
500 m
-1
-1.020
-980 m
3
2.91
3.09
1
-3
-3.09
-2.91
6
5.86
6.14
2
-6
-6.14
-5.86
10
9.80
10.20
5
-10
-10.20
-9.80
30
29.1
30.9
10
-30
-30.9
-29.1
60
58.6
61.4
20
-60
-61.4
-58.6
100
98.0
102.0
50
-100
-102.0
-98.0
300
291
309
100
-300
-309
-291
(12) Reduce calibrator output to minimum.
(13) Press TI keys as listed in (a) through (d) below.
SAVE.
(a)
F2 RECALL
(b)
(c)
Select CAL_A saved in 7. above using the
keys.
ENTER.
(d)
(14) Press TI keys as listed in (a) through (f) below.
(a) A, F4 INPUT A OPTIONS, using
20 MHz , ENTER.
keys select on A, ENTER and V ac,
(b) SCOPE, F2 READING, using
ENTER.
(c) CLEAR to exit softkey menu.
(d) s TIME ns, to select 10 mS/div.
(e) Using mV RANGE V to set the Input A sensitivity to 500mV/div.
(f) Using MOVE move the Input A ground level to the center grid line.
10