Figure 3. Temperature Adjustment Connection.

Calibration Procedure For Micro-Ohmmeter Model 6250 Page Navigation
  2    3    4    5    6  7
img
. . . .
TB 9-6625-135-40
(6) Press > to select adj L. Press > to enter the value of the resistance standard.
(7) Press the up key ^ to display MEASL. The low point is then adjusted.
(8) Press > to select adj H.
(9) Connect the decade resistor to the multimeter with a 4-wire connection. Adjust
the decade resistor (if necessary) to 115.54  as observed on the multimeter.
(10) Connect the decade resistor to the probe connector (fig. 3).
(11) Set the decade resistor to 15.54
= 40 C.
(12) Press > to enter the value of the resistance standard.
(13) Press the up key ^ to display MEASH. The high point is then adjusted. The
ADJ message indicates the adjustment is complete and was successful.
(14) Repeat paragraph 9 above.
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. If RTD temperature probe does not accompany TI, affix a DA Label 163 for
Temperature.
c. Annotate and affix DA label/form in accordance with TB 750-25.