TB 9-6625-149-50
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Calibration Data Card, DA Form 2416
a.
Forms, records. and
reports required for calibration personnel at all
levels are
prescribed by TM 38-750. DA Form 2416 must be annotated in accordance with TM 38-750
for each calibration performed.
b. Adjustments to be reported on DA Form 2416 are designated (R) at the end of the
sentence in which they appear. When adjustments are in tables, the (R) will follow the
designated adjustment. Report only those adjustments made and designated with (R).
3. Calibration Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description.
Test instrument parameters.
Performance specifications.
Power input requirements
117 or 234 vac 10%, 50 to 60 Hz
Sweep rates
.1 sec/cm to 5 sec/cm in 24 calibrated steps
Accuracy
3%
Sweep magnification
Increases the sweep amplifier gain 5 times
Accuracy
5%
Triggering facilities
Provides for manual control, preset stability
control, and fully automatic triggering.
Horizontal deflection system:
Deflection factor
Continuously variable from approximately .2
v/cm to 15 v/cm.
Frequency response
Dc to 240 kHz or better
Amplitude calibrator:
Output voltages
0.2 mv to 100 v p-p in 18 steps
Accuracy
3%
Current Probe Amplifier, Tektronix Types 131-1 through -4 with Current Probe, Tektronix Type P6016
Power requirements
115 vac 10 v, 50 to 60 Hz, 1 w
Sensitivity
1 ma/div to 1 amp/div in 10 calibrated steps
Accuracy
3%
50 Hz at 3 db down
Rise time
20 nsec
Maximum current rating
15 amps p-p
3