TB 9-6625-149-50
(3) TRIGGERING LEVEL control to 0 (zero).
(4) TRIGGERING MODE switch to AUTO.
(5) TRIGGER SLOPE switch to +INT.
(6) TIME/CM switch to .5 MILLISEC.
(7) TIME/CM VARIABLE control to CALIBRATED.
(8) STABILITY control to midrange.
(9) INTENSITY, FOCUS, and ASTIGMATISM controls for sharp trace.
(10) HORIZONTAL POSITION control to center the trace.
(11) AMPLITUDE CALIBRATOR switch to 100 VOLTS.
g. Turn
calibration adapter TEST FUNCTION switch to
HIGH LOAD.
h. Turn
calibration adapter VERTICAL POSITION control to
center trace on
crt.
SECTION IV
TEKTRONIX TYPE 541A
NOTE
Unless otherwise specified, verify the results of each test and
take corrective action whenever the test requirement is not
met before continuing with the calibration.
NOTE
When indications specified in paragraphs 8 through 18 are not
within tolerance, perform the power supply check prior to
making adjustments. After adjustments are made, repeat
if all other parameters are within tolerance.
6