TB 9-6625-149-50
11. Sweep Length and Registration
a. Performance Check
(1) Turn calibration adapter (A4) TEST FUNCTION switch to HIGH LOAD.
(2) Turn TI TRIGGERING MODE switch to AC and STABILITY control fully
clockwise. If sweep is not approximately 10.5 centimeters long, perform b(1) below.
(3) Connect time-mark generator (A10) to calibration adapter EXT INPUT, using
cable (B5).
(4) Set time-mark generator to .5 mS markers.
(5) Adjust TI STABILITY and TRIGGERING LEVEL controls for a stable
display.
(6) Turn TI HORIZONTAL DISPLAY switch to 5X MAG and position first time
marker behind center graticule line with HORIZONTAL POSITION control.
(7) Turn TI HORIZONTAL DISPLAY to NORM.
If marker does not remain
behind center graticule line, perform b(2) below.
b. Adjustments
(1) Adjust SWP LENGTH R176 (fig. 2) until sweep is approximately 10.5
centimeters long. (R)
(2) Adjust NORM. MAG. REGIS. R358 (fig. 2) until marker is centered behind
center graticule line. (R)
12. Triggering
a. Performance Check
(1) Connect TI CAL OUT to calibration adapter (A4) EXT INPUT, using cable (B5)
and, if necessary, adapter (B2).
(2) Position TI controls as listed in (a) through (c) below:
(a) AMPLITUDE CALIBRATOR switch to 5 VOLTS.
(b) TIME/CM switch to .5 MILLISEC.
(c) STABILITY control fully clockwise.
(3) Adjust calibration adapter VARIABLE control for 1 minor division of vertical
deflection on TI crt.
10