TB 9-6625-152-24
(8) Record measuring receiver indication.
(9) Use TI [ ], [ ] keys to highlight -1 dBr calibration point.
(10) Use TI rotary knob to alter data value until measuring receiver indicates 1 dB
less than indication recorded in (8) above.
(11) Use TI [ ], [ ] keys to step down to next calibration point in displayed table
(-2 dBr).
(12) Use TI rotary knob to alter data value until measuring receiver indicates 2 dB
less than indication recorded in (8) above.
(13) Repeat technique of (11) and (12) above for all remaining calibration points in TI
displayed table.
(14) Press TI [Return] key and then [Store Cal] key.
(15) Press TI [HELP SETUP], [Calibrate], [Output Level], [N out Ant in] keys.
(16) Connect measuring receiver to TI N type output connector and configure
measuring receiver for power measurement.
(17) Use TI [ ], [ ] keys to highlight 10MHz calibration point.
(18) Use TI rotary knob to alter data value until measuring receiver indicates
-21 dBm.
(19) Repeat technique of (17) and (18) above for all remaining frequencies in TI
displayed table.
(20) Press TI [N out N in] key.
(21) Use TI [ ], [ ] keys to highlight 10MHz calibration point.
(22) Use TI rotary knob to alter data value until measuring receiver indicates -21
dBm.
(23) Repeat technique of (21) and (22) above for all remaining frequencies in TI
displayed table.
(24) Press TI [BNC output] key.
(25) Connect measuring receiver to TI BNC output connector and configure
measuring receiver for power measurement.
(26) Use TI [ ], [ ] keys to highlight 10MHz calibration point.
(27) Use TI rotary knob to alter data value until measuring receiver indicates
+5 dBm.
(28) Repeat technique of (26) and (27) above for all remaining frequencies in TI
displayed table.
(29) Press TI [Return] key and then [Store Cal] key.
(30) Press TI [HELP SETUP], [Calibrate], [Attenuators] keys.
(31) Connect measuring receiver to TI N type output connector and configure
measuring receiver for power measurement.
(32) With TI 10 dB PAD CAL screen displayed and 10 MHz calibration point
highlighted, press [Pad In/Out] key to set pad state to OUT.
11