Figure 1. Dc Volts Output.
Figure 3. 4-Wire Connection.

Calibration Procedure For Range Calibrator, Models, 11683A And 11683A-H01 Page Navigation
  1    2    3  4  5  
img
. . . . . . .
TB 9-6625-158-24
(4) Adjust A2R1 (fig. 2) for 145.00 mV 2.00 mV dc. (R)
(5) Replace TI top panel.
Figure 2. Adjustment Location.
9. Resistance
a. Performance Check
(1) Set TI LINE OFF ON to OFF and FUNCTION switch to STANDBY.
(2) Connect equipment as shown in figure 3 below.
(3) Configure multimeter for 4-wire resistance measurement.
(4) Set TI RANGE switch to 300 W.
Multimeter will indicate within limits
specified in first row of table.
(5) Repeat technique of (4) above for remaining TI ranges listed in table 4.
Multimeter will indicate within limits specified in table 4.
5