Figure 2. Adjustment Location.

Calibration Procedure For Range Calibrator, Models, 11683A And 11683A-H01 Page Navigation
  1    2    3    4  5
img
. . . .
TB 9-6625-158-24
Figure 3. 4-Wire Connection.
Table 4. Resistance Check.
Test instrument
Multimeter indication
Limits (Ω)
Range
Min
Max
300 W
3143.3
3157.1
100 W
995.9
1000.2
30 W
315.14
316.52
10 W
99.749
100.18
3 W
31.58
31.718
b. Adjustments. None.
10. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.