TB 9-6625-173-24
a. Performance Check
(1) Connect equipment as shown in figure 1 below.
(2) Configure multimeter for dc volts measurement in auto range. Adjust the DC
power supply for a 12 V indication on the multimeter. Disconnect the multimeter used in
connection 1.
(3) Typical current drains should be ~0.6 mA during sleep mode, ~19 mA during a
1 Hz sample rate with RS232 communication (4.2 mA without RS232), and ~28 mA during
100 Hz sample rates with RS323 communication (17mA without RS232).
(4) Verify the Battery voltage register in the status table displays +12 V 0.5 V as
compared to the 12 V input.
(6) Verify the Panel Temp register in the status table displays the correct
temperature.
(7) Verify the Lithium Battery register in the status table reads greater than 3 V.
a. Performance Check
(1) Connect positive lead from calibrator to SE channel 1 and the negative lead to
SE channel 2. Also connect SE channel 2 to the nearest analog ground.
NOTE
Ensure "Public" is selected and use ports and flags to enable
the various flags for (True/False; or On/Off). A single click in
the `Radio Button' will enable and disable the desired flag. A
momentary wait is typical for the flag to engage/disengage.
(2) Download program Rangetst.CR1.
(3) Set Flag (1) true and set the calibrator for a +5000 mV output. Record the
reading to table 1 and verify specifications. If not within specifications perform b below.
Set the calibrator for a 0 V output and measure the offset voltage and record the reading to