• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Final Procedure
APPENDIX A. TEST PROGRAMS - Continued

Calibration Procedure For Data Logger, Model Cr1000 Page Navigation
  1    2    3    4    5  6  7    8    9  
TB 9-6625-173-24
APPENDIX A
TEST PROGRAMS
ADJUSTMENT PROGRAMS
Zero_DAC.CR1
--------------------------------Program #1-----------------------------
'Program: Zero_DAC.CR1
Const Range = mV2_5
Public Flag(8)
Public Batt
Public Diff1
BeginProg
Scan (1, sec,4,0)
Battery (Batt)
VoltDiff (Diff1,1,mV2_5,1,True,0,250,1.0,0)
ExciteV(Vx1,0,0)
AbortScan
NextScan
EndProg
Trim DAC reference
--------------------------------Program #2-----------------------------
'Program: 2V_DACadj.CR1
Const Range = mV2_5
Public Flag(8)
Public Batt
Public Diff1
BeginProg
Scan(1,sec,4,0)
Battery (Batt)
VoltDiff (Diff1,1,mV2_5,1,True,0,250,1.0,0)
ExciteV(Vx1,2000,0)
AbortScan
NextScan
EndProg
A-1


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.