• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-153. Calibration Description.
Table B-158. Resistance.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  54    55    56    57    58  59  60    61    62    63    64  
img
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Greenlee, Model CMI-100 Continued
Table B-154. DC Voltage.
Calibrator output
Test instrument
Limits (V dc)1
Applied
Range
(V dc)
(V dc)
Min
Max
360 m
400 m
357.1
362.9
3.6
4
3.571
3.629
36
40
35.71
36.29
120
400
118.9
121.1
280
400
277.7
282.3
360
400
357.1
362.9
-360
400
-362.9
-357.1
540
600
534
546
Allow 3 seconds for reading stabilization.
1
Table B-155. AC Voltage.
Calibrator output
Test instrument
Limits (V ac)1
Applied
Range
(V ac)
(V ac)
Min
Max
3.6 @ 40 Hz
4
3.552
3.648
3.6 @ 450 Hz
4
3.552
3.648
3.6 @ 60 Hz2
3.5003
3.7003
4
36 @ 40 Hz
40
35.52
36.48
36 @ 450 Hz
40
35.52
36.48
360 @ 40 Hz
400
355.2
364.8
360 @ 450 Hz
400
355.2
364.8
540 @ 40 Hz
600
527
553
540 @ 450 Hz
600
527
553
Allow 5 seconds for reading stabilization.
1
Press TI PEAK HOLD to On.
2
3 Set calibrator to STBY before verifying limits. Press TI PEAK HOLD to Off after limits verification.
Table B-156. DC Current (≤1000 A) (w/50 Turn Coil).
Calibrator output
Ammeter calibrator
Test instrument
Limits (A dc)1
Applied
Range
Range
(V dc)
(A dc)
(A dc)
Min
Max
0.72
20
400
352.3
367.7
1.8
20
1000
877
923
Allow 5 seconds for reading stabilization.
1
Table B-157. AC Current (≤1000 A) (w/50 Turn Coil).
Calibrator output
Ammeter calibrator
Test instrument
Limits (A ac)1
Applied
Range
Range
(V ac)
(A ac)
(A ac)
Min
Max
0.72 @ 40 Hz
20
400
351.8
368.2
0.72 @ 450 Hz
20
400
351.8
368.2
1.8 @ 40 Hz
20
1000
872
928
1.8 @ 450 Hz
20
1000
872
928
Allow 5 seconds for reading stabilization.
1
B-46


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.