Table B-170. Calibration Description.
Table B-174. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  59    60    61    62    63  64  65    66    67    68    69  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Megger, Model DET14C Continued
Table B-173. Resistance.
Resistance standard1
Test instrument
Limits (Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
0.5
AUTO
0.44
0.56
5
AUTO
4.8
5.2
50
AUTO
48.5
51.5
150
AUTO
141.5
158.5
300
AUTO
260
340
900
AUTO
810
990
1500
AUTO
975
2025
Substitute resistance standard for calibrator.