Table B-154. DC Voltage.
Table B-161. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  55    56    57    58    59  60  61    62    63    64    65  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Greenlee, Model CMI-100 Continued
Table B-158. Resistance.
Calibrator output
Test instrument
Limits (%/Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
1
+1.263% (192.4)1
190
400
-1.263% (187.6)
-1.158% (1.878 k)1
+1.158% (1.922 k)1
1.9 k
4k
-1.158% (18.78 k)1
+1.158% (19.22 k)1
19 k
40 k
-1.158% (187.8 k)1
+1.158% (192.2 k)1
190 k
400 k
-1.158% (1.878 M)1
+1.158% (1.922 M)1
1.9 M
4M
-1.684% (18.68 M)2
+1.684% (19.32 M)2
19 M
40 M
Allow 5 seconds for reading stabilization.
1
Allow 15 seconds for reading stabilization.
2
Table B-159. Frequency.
Function generator output
Test instrument
Applied
Range
Limits (Hz)
(Hz)
(Hz)
Min
Max
90 @ 100 mV
100
89.89
90.11
900 @ 100 mV
1000
898.9
901.1
9 k @ 100 mV
10 k
8.989 k
9.011 k
90 k @ 100 mV
100 k
89.89 k
90.11 k
720 k @ 1 V
800 k
719 k
721 k
Table B-160. Frequency (≤500 A) (w/Single Conductor).
Calibrator output
Test instrument
Applied
Range
Limits (Hz)
(Hz)
(Hz)
Min
Max
90 @ 5 A
100
89.50
90.50
900 @ 5 A
1000
895.0
905.0
B-47