• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-239. DC Current (=500 A) (w/50 Turn Coil).Calibrator output
Table B-247. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  78    79    80    81    82  83  84    85    86    87    88  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Extech, Model MA640 Continued
Table B-245. Capacitance.
Capacitance standard
Test instrument
Applied
Range
Limits (F)
(F)
(F)
Min
Max
36 n
40 n
34.36 n
37.64 n
360 n
400 n
348.7 n
371.3 n
3.6 μ
4μ
3.487 μ
3.713 μ
36 μ
40 μ
34.87 μ
37.13 μ
360 μ
400 μ
344.6 μ
375.4 μ
3600 μ
3410 μ
3790 μ
4000
Table B-246. Temperature.
Cool/Heat source
Test instrument
Limits
Applied
Range
Min
Max
0 F
F
-9 F
9 F
1000 F
F
991 F
1039 F
0 C
C
-5 C
5 C
600 C
C
577 C
623 C