Table B-247. Calibration Description.
Table B-256. DC Voltage.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  80    81    82    83    84  85  86    87    88    89    90  
img
.
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Fieldpiece, Model SC66 Continued
Table B-251. AC Current (≤500 A) (w/50 Turn Coil).
Calibrator output
Test instrument
Applied
Range
Limits (A dc)
(A ac)
(A ac)
Min
Max
0.36 @ 50 Hz
20
17.41
18.59
0.36 @ 60 Hz
20
17.41
18.59
3.6 @ 50 Hz
200
174.1
185.9
3.6 @ 60 Hz
200
174.1
185.9
3.6 @ 400 Hz
200
170.5
189.5
Table B-252. Resistance.
Calibrator output
Test instrument
Limits (Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
190
200
-1.158% (187.8)
+1.158% (192.2)
190 k
200 k
-1.158% (187.8 k)
+1.158% (192.2 k)
Table B-253. Capacitance.
Capacitance standard
Test instrument
Applied
Range
Limits (F)
(F)
(F)
Min
Max
180 μ
200 μ
174.1
185.9
Table B-254. Temperature.
Cool/Heat source
Test instrument
Limits
Applied
Range
Min
Max
10 F
F
9 F
11 F
190 F
F
189 F
191 F
B-72