• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table B-266. DC Current (=500 A).Calibrator output
Table B-275. Calibration Description.

Calibration Procedure For Digital Clamp Meters (General) Page Navigation
  85    86    87    88    89  90  91    92    93    94  
img
. .
TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
TIF Instruments, Model TIF1000DC
Table B-271. Calibration Description.
Test instrument parameters
Performance specifications
DC VOLTAGE
Range:0 to 200 V dc
Accuracy: (1.0% rdg + 1 dgt)
DC CURRENT
Range: 0 to 1200 A dc
Accuracy: (1.0% rdg + 1 dgt) 0 to 800 A range;
+0% to -5% rdg 800 to 1200 A range
Range: 0 to 2000 Ω
RESISTANCE
Accuracy: (3.0% rdg + 1 dgt)
Table B-272. DC Voltage.
Calibrator output
Test instrument
Applied
Range
Limits (V dc)
(V dc)
(V dc)
Min
Max
60
200
59.3
60.7
140
200
138.5
141.5
180
200
178.1
181.9
-180
200
-181.9
-178.1
Table B-273. DC Current (≤1000 A) (w/50 Turn Coil).
Calibrator output
Ammeter
Test instrument
calibrator
Applied
Range
Range
Limits (A dc)
(V dc)
(A dc)
(A dc)
Min
Max
0.36
20
200
178.1
181.9
2.0
20
1200
950
1000
Table B-274. Resistance.
Calibrator output
Test instrument
Limits (Ω)
Applied
Range
(Ω)
(Ω)
Min
Max
1.9 k
2000
-3.05% (1842)
+3.05% (1958)
B-77


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.