TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Fluke, Model 376 Continued
Table B-287. iFlex Current Probe Accuracy (with Simulation). 1
Calibrator
Test instrument
Output
Min (Amps)
Max (Amps)
3 mV @ 50 Hz
96.5
103.5
30 mV @ 50 Hz
965
1035
60 mV @ 50 Hz
1935
2065
75 mV @ 50 Hz
2420
2580
750 mV @ 500 Hz
2420
2580
For the iFlex simulated tests, the "Test Lead w/retractable sheath" lead and "Test Probe for iFlex"
1
probe are required. They are used to connect the calibrator output to the iFlex input. Calibrator
Output HI goes to the iFlex 2 mm jack (on the far left) and the Calibrator Output LO goes to the black
COM jack.
Table B-288. iFlex Current Probe Accuracy (with 15-Turn Coil).
Current
Calibrator
Test instrument
Calibrator
Coil
Output to TI
Turn(s)
Output
Range
Min (Amps)
Max (Amps)
10 A @ 50 Hz
---------------
1
10
9.2
10.8
0.332 V @ 50 Hz
100 A
15
498
484.5
515.5
0.6 V @ 50 Hz
100 A
15
900
587.5
927
2 V @ 440 Hz
20 A
15
300
290.5
309.5
B-81/(B-82 blank)