TB 9-6625-174-24
APPENDIX B
PERFORMANCE TABLES
Fluke, Model 376 Continued
Test instrument
Standard
Output (Ω)
Min (Ω)
Max (Ω)
Range
60
58.9
61.1
300
296.5
303.5
540
534.1
545.9
Auto
3000
2965
3035
5400
5341
5459
30 k
29650
30350
54 k
53410
54590
Test instrument
Min
Max
10 F
9.5 F
10.5 F
500 F
491.0 F
509.0 F
900 F
887.0 F
913.0 F
Table B-284. Ac Current Accuracy.
Current
Test instrument
Calibrator
Calibrator
Coil
Output to TI
Turn(s)
Output
Range
Min (Amps)
Max (Amps)
10 A @ 50 Hz
---------------
1
10
9.3
10.7
0.332 V @ 50 Hz
100 A
15
498
489.5
510.5
0.6 V @ 50 Hz
100 A
15
900
881.5
918.5
2 V @ 440 Hz
20 A
15
300
292
308
Table B-285. Dc Current Accuracy.
Current
Test instrument
Calibrator
Calibrator
Coil
Output to TI
Turn(s)
Output
Range
Min (Amps)
Max (Amps)
10 A
---------------
1
10
9.3
10.7
0.332 V 1
100 A
15
498
489.5
510.5
0.6 V
100 A
15
900
881.5
918.5
1 Connect
the calibrator to the Ammeter calibrator with the 15 turn current coil.
Calibrator
Test instrument
Range
Output (Hz)/(Amps)
Min (Hz)
Max (Hz)
(A ac)
39.3
40.7
40 @ 5 A rms
60 @ 5 A rms
59.2
60.8
Auto
90 @ 5 A rms
89.0
91.0
450 @ 10 A rms
447.2
452.8
B-80