TB 9-6625-1866-50
(9) Turn VOLTS RANGE switch to 500 and repeat a(5) and (6) above.
10. 50-Volt Range
a. Performance Check
(1) Turn, VOLTS RANGE switch to 50 and adjust dc voltage standard (A2, table 1)
for full-scale indication on UUT meter.
(2) Turn NULL switch to 10, 1, and .1 while adjusting dc voltage standard for 0
(zero) indication on UUT meter. Dc voltage standard will indicate between 49.975 and
50.025 volts. If not, perform b below.
(3) Turn NULL switch to VTVM.
(4) Adjust dc voltage standard for 0 (zero) indication on UUT meter.
b. Adjustments
(1) Turn NULL switch to VTVM.
(2) Adjust dc voltage standard for 50.000 volts.
(3) Turn NULL switch to 10, 1, and .1 while adjusting P3 (fig. 1) for 0 (zero)
indication on UUT meter.
11. 5-Volt Range
a. Performance Check
(1) Turn UUT VOLTS RANGE switch to 5.
(2) Adjust dc voltage standard (A2, table 1) for full-scale indication on UUT meter.
(3) Turn NULL switch to 10, 1, .1 and .01 while adjusting dc voltage standard for 0
indication on UUT meter. Dc voltage standard will indicate between 4.9975 and 5.0025
volts. If not, perform b below.
(4) Turn NULL switch to VTVM.
(5) Adjust dc voltage standard for 0 (zero) indication on UUT meter.
b. Adjustments
(1) Turn NULL switch to VTVM.
(2) Adjust dc voltage standard for 5.00000 volts.