TB 9-6625-1866-50
AN/USM-98A and AN/ USM-98B (John Fluke Models 801B, 801BR, and 80 1B/AG) will be
c. Equipment and Accessory Identification. The equipment and accessories
referred to throughout this bulletin are identified in tables 1, 2, 6 and 7 and section IV.
d. Equipment Setup. Disconnect instructions are not contained in this bulletin.
5. Differences Among Models. Differential Voltmeter 801R is a
rack-mounted version
of the 801. AN/USM-98 is the military designation for the 801. Differential Voltmeter
801B contains a standard cell; the 801B/AG contains a Zener diode. The 801BR is the rack-
mounted version of the 801B. AN/USM-98A is the military designation for the 801B and
801B/AG. The AN/USM-98B is similar to the 801B. Front panel designations may vary
among models.
SECTION II
AN/USM-98 (J-F 801 AND 801R)
6. Equipment and Accessory Required. Table 1
lists minimum
use specifications of
equipment required for calibration performance checks and adjustments. Table 2 lists the
required accessory. Tables 1 and 2 are provided to assist in the selection of required
equipment and accessory. For specific item identification refer to section IV.
Table 1. Minimum Specifications of Equipment Required
Item
Number
Common Name
Minimum Use Specifications
A1
RANGE: 105 to 125 vac
ACCURACY: 1%
A2
DC VOLTAGE STANDARD
RANGE: 0 to 520 v
ACCURACY: 0.016%
Table 2. Accessory Required
Item
Number
Common Name
Description
B1
CABLE
30-in., RG-58/U; double banana plug
terminations
7. Preliminary Procedure
NOTE
Personnel should familiarize themselves with the entire
a. With no power applied to UUT, mechanically zero meter, using meter mechanical
zero adjustment screw.
4