TB 9-6625-1914-24
a. Performance Check
(2) Position TI controls as listed in (a) through (g) below:
(a) RANGE MHz switch to 0-110 (not on all models).
(c) SCAN WIDTH (red) switch to PER DIVISION.
(d) SCAN WIDTH PER DIVISION switch to 20 kHz.
(g) TUNING STABILIZER switch to ON (up).
Figure 9. Spurious response - equipment setup.
(3) Position IF plug-in switches as listed in (a) through (e) below:
(a) SCAN TIME PER DIVISION to 2 ms.
(b) LOG REF LEVEL to -40 dBm.
(c) VIDEO FILTER to OFF.
(d) SCAN MODE to INT.
(e) SCAN TRIGGER to AUTO.