Tuning Dial Accuracy
Figure 11. Test instrument - adjustment locations.

Calibration Procedure For Spectrum Analyzer, If, Lf, And Rf Plug-In Units, Models 8552, 8553, 8554, 8555, And 8556 Page Navigation
  30    31    32    33    34  35  36    37    38    39    40  
img
. . . . .
TB 9-6625-1914-24
(2) Position TI controls as shown in (a) through (e) below:
(a) BANDWIDTH switch to 30 kHz.
(b) SCAN WIDTH (red) switch to PER DIVISION.
(c) SCAN WIDTH PER DIVISION switch to 10 MHz.
(d) TUNING STABILIZER switch to OFF.
(e) FREQUENCY FINE TUNE control centered.
(3) Position FREQUENCY control and signal generator frequency controls as
listed in table 8. If signal pip display is not within 1 division of center graticule line for
each setting, perform b below.
Table 8. Tuning Dial Accuracy
Signal generator 1
FREQUENCY
(MHz)
(MHz)
200
200
400
400
600
600
800
800
1000
1000
1
Set signal generator amplitude for usable display.
b. Adjustments
(1) Connect frequency counter to 1500 MHz A8J1 (fig. 11), using cable (Hewlett-
Packard, Model 115926001, part of service kit). NOTCH FILTER FL2 assembly (fig. 11)
must be disconnected and removed for this connection.
(2) Adjust 2ND LO ADJ (fig. 11) as necessary to obtain frequency counter indication
between 1499.9 and 1500.1 MHz (R).
(3) Disconnect frequency counter and reinstall NOTCH FILTER FL2 assembly
removed in (1) above.
(4) Set BANDWIDTH switch to 100 kHz and SCAN WIDTH PER DIVISION
switch to 20 MHz.
(5) Set IF plug-in SCAN TRIGGER switch to LINE.
(6) Adjust FREQUENCY control to 0 MHz.
35