TB 9-6625-1914-35
a. Performance Check
(2) Position TI controls as listed in (a) through (d) below:
(a) FREQUENCY control to 0 MHz.
(c) SCAN WIDTH (red) switch to 0-1250 MHz.
(3) Set IF plug-in LOG/LINEAR switch to LINEAR and LINEAR SENSITIVITY
(LOG REF LEVEL) switch to 1 mV/DIV.
(4) Adjust signal generator frequency to 100 MHz and output level to produce a
7-division signal pip display. Note output level of signal generator.
(5) Slowly increase signal generator frequency to 500 MHz while maintaining
output level noted in (4) above. Note frequency which produces highest signal pip
amplitude.
(6) Adjust signal generator output level to produce an 8-division signal pip display