TB 9-6625-1914-35
HEWLETT-PACKARD, MODEL 8555( )
33. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 33 and
34 are preparatory to the
calibration
process. Personnel should become familiar with the applicable section before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
NOTE
Unless otherwise specified, verify the result of each test and,
whenever the test requirement is not met, take corrective
action before continuing with the calibration. Adjustments
required to calibrate the TI are included in this procedure.
Additional maintenance information is contained in the
manufacturer's manual for this TI.
NOTE
Unless otherwise specified, all controls and control settings
refer to the TI.
a. Mate TI with IF plug-in and install in display section.
c. Connect autotransformer to a
115 V ac
source and
adjust for a
115 V ac
output.
d. Energize equipment and allow 1 hour for equipment to warm-up and stabilize.
e. Position TI controls as listed in (1) through (9) below:
(1) BAND switch to .01-2.05 GHz.
(3) FINE TUNE control centered.