Figure 1. Test instrument - right side view.
Figure 3. Wattmeters TS-779B/U and Model 451 BOLO RES switch S1 test point location - right side view

Calibration Procedure For Wattmeters An/Urm-98, Ts-770/U Models 430C And 430Cr; And Model 451: Bolometer 9Thermistor Mounts Mx-2114/U, And Models 477 Page Navigation
  1    2    3    4    5  6  7    8    9    10  
img
. . . . . .
TB 9-6625-1918-24
(3) Adjust ZERO SET COARSE and FINE controls for a 0.190 Vac indication
on multimeter.
(4) Adjust R63 FS ADJ (fig. 1) for a full-scale indication on TI meter (R).
9. Bridge Balance
a. Performance Check
(1) Connect multimeter between TEST POINT 1 (fig. 2) and chassis ground. For
TS-779B/U and General Microwave Model 451, connect multimeter to junction of R2 and S1
and terminals 4 and 5 (fig. 3).
Figure 2. Test instrument - top view.
(2) Set BOLO BIAS CURRENT (BIAS CURRENT on TS-779/U) switch and
adjust ZERO SET COARSE and FINE controls for a 3.0 V ac indication on multimeter.
(3) Connect multimeter between bolometer and TI BOLOMETER connector.
If
multimeter does not indicate between 0.98 and 1.02 V ac, perform b below.
6