APPENDIX A. TEST INSTRUMENT SPECIFICATIONS - Continued
APPENDIX B. CONVERSIONS OF VSWR TO RETURN LOSS DB

Calibration Procedure For Power Sensors And Thermistor Mounts 10 Mhz To 40 Ghz (General) Page Navigation
  18    19    20    21    22  23  24    25  
img
TB 9-6625-1932-24
APPENDIX A
TEST INSTRUMENT SPECIFICATIONS
Power Sensors - Continued
Standard limitations
from test report
indications
VSWR
(Kb factor)
max
Power range
Frequency
Manufacturer
Model
range
1 W to 100 mW
8485A3
50 to 100 MHz: 1.15
Hewlett-Packard 50 MHz to
50 MHz to 2 GHz:
100 MHz to 2 GHz: 1.10
26.5 GHz  (-30 dBm to +20 dBm)
3% of calibration
factor listed on
2 to 12.4 GHz: 1.15
mount
12.4 to 18 GHz: 1.20
50 MHz to 10 GHz:
18 to 26.5 GHz: 1.25
3%
14 GHz:  3.2%
18 GHz:  3.6%
22 GHz:  3.7%
26.5 GHz: 4.0%
1Hewlett-Packard,
Model 11708A (30 dB) attenuator must be calibrated with power sensor.
2Not
checked below 10 MHz.
3Weinschel, Model 991830 dB attenuator is needed in the calibration of Hewlett-Packard, Model 8484A (attenuator
connection factor will be considered when calculating power level).
A-4