Figure 8. Calibration factor equipment setup.
Final Procedure

Calibration Procedure For Power Sensors And Thermistor Mounts 10 Mhz To 40 Ghz (General) Page Navigation
  13    14    15    16    17  18  19    20    21    22    23  
img
. . .
TB 9-6625-1932-24
12. Calibration Factors
a. Performance Check
(1) Press synthesized signal generator RF output pushbutton to OFF position.
(2) Adjust synthesized signal generator frequency controls for first cardinal point
determined in 11 a above.
(3) Set power meter No. 2 RANGE switch to COARSE ZERO and adjust front
panel COARSE ZERO control for a zero meter indication.
(4) Ensure multimeter input terminals are isolated from chassis ground for (5) below.
(5) Connect multimeter positive lead to power meter No. 2 VCOMP and connect
multimeter negative lead to power meter VRF.
(6) Set power meter No. 2 RANGE switch to 1 mW.
(7) Adjust multimeter (dc mode) controls to measure microvolts.
(8) Press and hold power meter No. 2 FINE ZERO control and record multimeter
indication as V0.
(9) Release power meter No. 2 FINE ZERO control.
(10) Set synthesized signal generator RF output pushbutton to ON position.
(11) Adjust synthesized signal generator output controls for a power meter No. 2
indication (reference) of 1 mW full scale or suitable power level full scale indication within
range of the TI.
(12) Record multimeter indication as V1.
(13) Disconnect multimeter negative lead from power meter No. 2 VRF and connect
multimeter negative lead to power meter No. 2 chassis ground.  Record multimeter
indication as Vcomp.
(14) Calculate and record the power level from the below listed formula as P1:
2 VCOMP (V1 - V0) + V02 - V12
P1 =
4 R (calibration factor)
P1 = Output power level
VCOMP = Value recorded in (13) above.
V1 =Value recorded in (12) above
V0 = Value recorded in (8) above
R = Value recorded in 11 f above
Calibration Factor = Value for standard thermistor mount at test frequency
(15) Set power meter No. 1 RANGE switch to COARSE ZERO and adjust front
panel COARSE ZERO control for a zero meter indication.
(16) Set power meter No. 1 RANGE switch to 1 mW and zero power meter No. 1.
(17) Disconnect power meter No. 2 from synthesized signal generator and connect
equipment as shown in figure 8, CONNECTION B.